This site is targeted at medical and radiology professionals, contains user contributed content, and material that may be confusing to a lay audience. Use of this site implies acceptance of our Terms of Use.

1.5T vs 3T

Comparing 1.5T vs 3T MRI systems identifies in a number of differences:

Signal to noise ratio

Theoretically signal increases proportional to the square of the static field strength whereas noise increases linearly. This implies that in a perfect system SNR of a 3.0T system would be twice as good as at 1.5T. In reality, due to increase in susceptibility effects in most tissues the actual improvement is only in the 30-60% range (instead of 100%). With this increased SNR, the spatial resolution and/or acquisition time can be improved depending on which is more important for the particular case.

Specific absorption rate (SAR)

SAR is defined as the amount of radiofrequency energy (joules) deposited tissues (kg). The limit set by the FDA is an amount which results in increase of 1 degree centigrade in any tissue 2. SAR is proportional to the static field (B0) to the power of 2, meaning a 3.0T system and 4 times as much as a 1.5T system. Additionally SAR is proportional to:

  • B0 squared
  • pulse duration and length
  • pulse number
  • slice number
  • flip angle

The last of the above, results in standard spin echo sequences which have 90 degree flips depositing to much energy. As a result there is a greater use of gradient echo sequences. Unfortunately these are imaging T2* not T2 and therefore are more susceptible to local field artefacts. Largely, these problems have been overcome with modern units.

Acoustic noise

Rapid gradient switching leads to increase in the intensity of the acoustic noise, which has required better insulation of both the unit itself and the room.


Related articles

MRI physics

Updating… Please wait.
Loadinganimation

Alert_accept

Error Unable to process the form. Check for errors and try again.

Alert_accept Thank you for updating your details.