Crosstalk artifact
Case contributed by:
Dr Roberto Schubert
From the case:
Crosstalk artifact
Modality:
MRI
Case Discussion:
Also known as slice overlap artifact. If multislice acquisitions are overlapping, the protons belonging to more than one slice get saturated.
Prevention: use more shallow angles or move the overlapping region dorsally, out of the anatomy of interest.