The anode represents the component of the x-ray tube where x-rays is produced. It is a piece of metal, shaped in the form of a bevelled disk with the diameter between 55 and 100mm, 7mm thick, connected to the positive side of the electrical circuit. The anode converts the energy of the electrons into X-rays and dissipates the heat, considered the byproduct.

Most X-ray tubes are built of tungsten (Z=74). Tungsten has the high atomic number Z and has a high melting point of 3370°C with the low rate of evaporation. Alloy containing tungsten and rhenium is also used, since 5-10% of rhenium prevents crazing of the anode surface. The body of the anode is made of the materials that are light and have a good heat storage capacity, like molybdenum and graphite. Molybdenum is also often used as the surface material for the anodes used in mammography, for its characteristics: intermediate atomic number – with the produced X-rays of the energies suited for this purpose. Some of the anodes used for mammography are also made of rhodium (Z=45), producing more penetrating radiation, preferably for use in dense breast imaging.

Anodes are designed as beveled disks attached to the large copper rotor of the electric motor, rotating them at the speeds up to 10,000 RPM, with the temperature of 2000°C. The purpose of the rotation is to dissipate the heat. Most rotating anodes actually represent rather complex electromechanical systems consisting of approximately 350 pieces, taking around 150 assembly operations.

The anode disc rotates and is subjected to a focused beam of electrons emanated from the cathode, which are then accelerated by the high potential difference between the cathode and the target disc. When the electron beam hit the target, i.e. the focal spot, it then produces the X-ray beam. The anode angle is measured from vertical to the angle of the anode and is about 15°. A smaller angle would make a smaller focal spot.

The whole anode is not included in the X-ray production. The X-rays are produced on the rather small rectangular surface – the focal spot. Some X-rays have two focal spots, chosen according to the imaging procedure.

Share article

Article information

rID: 8178
Section: Physics
Synonyms or Alternate Spellings:

Support Radiopaedia and see fewer ads

Cases and figures

  • Drag
    Figure 1: X-ray tube diagram
    Drag here to reorder.
  • Updating… Please wait.
    Loadinganimation

    Alert accept

    Error Unable to process the form. Check for errors and try again.

    Alert accept Thank you for updating your details.