Beam width artifact

Changed by Bruno Di Muzio, 26 Oct 2014

Updates to Article Attributes

Body was changed:

Ultrasound beam width artifact occurs when a highly reflective object located beyond the widened ultrasound beam (after, after the focal zone) may create false, creates false detectable echoes that are displayed as overlapping the structure of interest.

To understand this artifact, it is important to remember that the ultrasound beam is not uniform with depth, the main beam leaves the transducer with the same width as it, then narrows as it approaches the focal zone and widens again distal to this zone 1

Usually, it occurs when scanning an anechoic structure and some peripheral echoes are identified, i.e. gas bubbles in the duodenum simulating small gallstones and peripheric echoes in the bladder. 

It is possible to avoid this artifact adjusting the focal zone to the depth level of interest and by placing the transducer at the center of the object being studied. 

  • -<![endif]--><!--StartFragment--><strong>Ultrasound</strong> <strong>beam width artifact</strong> occurs when a highly reflective object located beyond the widened ultrasound beam (after focal zone) may create false detectable echoes.</p><p>To understand this artifact, it is important to remember that the ultrasound beam is not uniform with depth, the main beam leaves the transducer with the same width as it, then narrows as it approaches the <a title="focal zone" href="/articles/focal-zone">focal zone</a> and widens again distal to this zone<sup> 1</sup>. </p><p>Usually, it occurs when scanning an anechoic structure and some peripheral echoes are identified, i.e. gas bubbles in the duodenum simulating small gallstones and peripheric echoes in the bladder. </p><p>It is possible to avoid this artifact adjusting the focal zone to the depth level of interest and by placing the transducer at the center of the object being studied. </p><p><!--EndFragment--></p><p><!--EndFragment--></p>
  • +<![endif]--><!--StartFragment--><strong>Ultrasound</strong> <strong>beam width artifact</strong> occurs when a reflective object located beyond the widened ultrasound beam, after the focal zone, creates false detectable echoes that are displayed as overlapping the structure of interest.</p><p>To understand this artifact, it is important to remember that the ultrasound beam is not uniform with depth, the main beam leaves the transducer with the same width as it, then narrows as it approaches the <a href="/articles/focal-zone">focal zone</a> and widens again distal to this zone<sup> 1</sup>. </p><p>Usually, it occurs when scanning an anechoic structure and some peripheral echoes are identified, i.e. gas bubbles in the duodenum simulating small gallstones and peripheric echoes in the bladder. </p><p>It is possible to avoid this artifact adjusting the focal zone to the depth level of interest and by placing the transducer at the center of the object being studied. </p><p><!--EndFragment--></p><p><!--EndFragment--></p>

References changed:

  • 1. Feldman MK, Katyal S, Blackwood MS. US artifacts. Radiographics. 2009;29 (4): 1179-89. <a href="http://dx.doi.org/10.1148/rg.294085199">doi:10.1148/rg.294085199</a> - <a href="http://www.ncbi.nlm.nih.gov/pubmed/19605664">Pubmed citation</a><span class="auto"></span>

Tags changed:

  • ultrasound

Sections changed:

  • Imaging Technology