MRI physics
The physics of MRI are complicated and much harder to understand than those underpinning image generation in plain radiography, CT or ultrasound.
What follows is a very abbreviated, 'broad strokes' description of the process. Essentially, the process can be broken down into four parts:
- preparation
- excitation
- spatial encoding
- signal acquisition
For a more detailed description of each part of the process, please refer to the links scattered throughout this introduction and at the bottom of the page.
Preparation
The patient is placed in a static magnetic field produced by the magnet of the MRI scanner. In living tissues there are a lot of hydrogen atoms included in water molecules or in many different other molecules. The proton, the nucleus of hydrogen, possesses an intrinsic magnetization called spin. The spin magnetization vector precesses (rotates) around the magnetic field at a frequency called the Larmor frequency, which is proportional to the magnetic field intensity. The resulting magnetization of all protons inside the tissues aligns parallel to the magnetic field. The parallel magnetization scales with the magnetic field intensity, basically at 3 T it will be twice the value obtained at 1.5 T. Additional preparation sequences can also be performed to manipulate the magnetization and so the image contrast, e.g. inversion preparation.
Excitation
During the image acquisition process, a radiofrequency (RF) pulse is emitted from the scanner. When tuned to the Larmor frequency, the RF pulse is at resonance: it creates a phase coherence in the precession of all the proton spins. The duration of the RF pulse is chosen such that it tilts the spin magnetization perpendicularly to the magnetic field. When a receiving coil (an electrical conductor) is put in the vicinity of the tissue, the transverse magnetization, that still rotates as the Larmor precession, will generate an electric current in the coil by Faraday induction: this is the nuclear magnetic resonance (NMR) signal.
The NMR signal is attenuated due to two simultaneous relaxation processes. The loss of coherence of the spin system attenuates the NMR signal with a time constant called the transverse relaxation time (T2). Concurrently, the magnetization vector slowly relaxes towards its equilibrium orientation that is parallel to the magnetic field: this occurs with a time constant called the spin-lattice relaxation time (T1). The contrast in MR images originates from the fact that different tissues have, in general, different T1 and T2 relaxation times; as this is especially true for soft tissues, it explains the excellent soft tissue contrast of MRI.
Spatial encoding
Spatial encoding of the MRI signal is accomplished through the use of magnetic field gradients (smaller additional magnetic fields with an intensity that linearly depends on their spatial location): spins from protons in different locations precess at slightly different rates. The portion of the gradient coils and the associated current that is perpendicular to the main magnetic field cause a force (Lorentz force) on the coils. The gradients are turned on and off very quickly in this process causing them to vibrate and producing the majority of the acoustic noise during MR image acquisition.
Signal acquisition
When using magnetic field gradients, the obtained NMR signal contains different frequencies corresponding to the different tissue spin positions and is called the MRI signal. After sampling, the analog MRI signal is digitized and stored for processing, which consists of a separation of the signal contributions from different spatial locations represented by pixels in the final image. This is achieved by a mathematical operation called a Fourier transform.
Standard exam
Multiple image sets are obtained in a standard examination protocol (which varies from facility to facility). Exam times vary according to the part of the anatomy being studied, pathology expected, radiologist preferences, and the scanner hardware and software used. Occasionally, a contrast medium may be used to enhance images, this will also usually prolong the scan time. Typically, exams are ordered without and with contrast for comparison purposes. Very rarely, and only in certain circumstances are exams ordered with contrast only. After the MRI exam the patient is removed from the scanner and given post-procedure instructions (information about contrast medium and/or sedation if used).
Related Radiopaedia articles
Imaging physics
- imaging physics
- imaging in practice
- imaging technology
-
x-ray physics
- ionizing radiation
- interaction with matter
- x-ray spectrum
- radiation units
- effective dose
- exposure
- legacy units
- radiation safety
- radiation damage (biomolecular)
- radiation damage (skin injury)
- stochastic effect
- CT physics
-
MRI physics
- B0
- chemical shift
- dependence of magnetization (proton density, field strength and temperature)
- echo time
- eddy currents
- electromagnetic induction
- Ernst angle
- flip angle
- Larmor frequency
- magnetic dipole magnetic field gradient
- magnetic susceptibility
- magnetism
- molecular tumbling rate effects on T1 and T2
- net magnetization vector (NMV)
- relaxation
- repetition time
- resonance and radiofrequency (RF)
- units of magnetism
- ultrasound physics
- nuclear medicine physics
Imaging technology
- imaging technology
- imaging physics
- imaging in practice
-
x-ray
- x-ray physics
- x-ray in practice
- x-ray production
- x-ray tubes
- tube rating
- filters
- automatic exposure control (AEC)
- beam collimators
- grids
- air gap technique
- cassette
- intensifying screen
- x-ray film
- image intensifier
- digital radiography
- digital image
- mammography
- x-ray artifacts
- radiation units
- radiation safety
- radiation detectors
- fluoroscopy
-
computed tomography (CT)
- CT physics
- CT in practice
- CT technology
- CT image reconstruction
- CT image quality
- CT dose
- CT contrast media
-
CT artifacts
- patient-based artifacts
- physics-based artifacts
- hardware-based artifacts
- ring artifact
- tube arcing
- out of field artifact
- air bubble artifact
- helical and multichannel artifacts
- CT safety
- history of CT
-
MRI
- MRI physics
- MRI in practice
- MRI hardware
- signal processing
-
MRI pulse sequences (basics | abbreviations | parameters)
- T1 weighted image
- T2 weighted image
- proton density weighted image
- chemical exchange saturation transfer
- CSF flow studies
- diffusion weighted imaging (DWI)
- echo-planar pulse sequences
- fat-suppressed imaging sequences
- gradient echo sequences
- inversion recovery sequences
- metal artifact reduction sequence (MARS)
-
perfusion-weighted imaging
- techniques
- derived values
- saturation recovery sequences
- spin echo sequences
- spiral pulse sequences
- susceptibility-weighted imaging (SWI)
- T1 rho
- MR angiography (and venography)
-
MR spectroscopy (MRS)
- 2-hydroxyglutarate peak: resonates at 2.25 ppm
- alanine peak: resonates at 1.48 ppm
- choline peak: resonates at 3.2 ppm
- citrate peak: resonates at 2.6 ppm
- creatine peak: resonates at 3.0 ppm
- functional MRI (fMRI)
- gamma-aminobutyric acid (GABA) peak: resonates at 2.2-2.4 ppm
- glutamine-glutamate peak: resonates at 2.2-2.4 ppm
- Hunter's angle
- lactate peak: resonates at 1.3 ppm
- lipids peak: resonates at 1.3 ppm
- myoinositol peak: resonates at 3.5 ppm
- MR fingerprinting
- N-acetylaspartate (NAA) peak: resonates at 2.0 ppm
- propylene glycol peak: resonates at 1.13 ppm
-
MRI artifacts
- MRI hardware and room shielding
- MRI software
- patient and physiologic motion
- tissue heterogeneity and foreign bodies
- Fourier transform and Nyquist sampling theorem
- MRI contrast agents
- MRI safety
-
ultrasound
- ultrasound physics
-
transducers
- linear array
- convex array
- phased array
- frame averaging (frame persistence)
- ultrasound image resolution
- imaging modes and display
- pulse-echo imaging
- real-time imaging
-
Doppler imaging
- Doppler effect
- color Doppler
- power Doppler
- B flow
- color box
- Doppler angle
- pulse repetition frequency and scale
- wall filter
- color write priority
- packet size (dwell time)
- peak systolic velocity
- end-diastolic velocity
- resistive index
- pulsatility index
- Reynolds number
- panoramic imaging
- compound imaging
- harmonic imaging
- elastography
- scanning modes
- 2D ultrasound
- 3D ultrasound
- 4D ultrasound
- M-mode
-
ultrasound artifacts
- acoustic shadowing
- acoustic enhancement
- beam width artifact
- reverberation artifact
- ring down artifact
- mirror image artifact
- side lobe artifact
- speckle artifact
- speed displacement artifact
- refraction artifact
- multipath artifact
- anisotropy
- electrical interference artifact
- hardware-related artifacts
- Doppler artifacts
- aliasing
- tissue vibration
- spectral broadening
- blooming
- motion (flash) artifact
- twinkling artifact
- acoustic streaming
- biological effects of ultrasound
- history of ultrasound
-
nuclear medicine
- nuclear medicine physics
- detectors
- tissue to background ratio
-
radiopharmaceuticals
- fundamentals of radiopharmaceuticals
- radiopharmaceutical labeling
- radiopharmaceutical production
- nuclear reactor produced radionuclides
- cyclotron produced radionuclides
- radiation detection
- dosimetry
- specific agents
- carbon-11
- chromium-51
- fluorine agents
- gallium agents
- Ga-67 citrate
- Ga-68
- iodine agents
-
I-123
- I-123 iodide
- I-123 ioflupane (DaTSCAN)
- I-123 ortho-iodohippurate
- I-131
-
MIBG scans
- I-123 MIBG
- I-131 MIBG
-
I-123
- indium agents
- In-111 Octreoscan
- In-111 OncoScint
- In-111 Prostascint
- In-111 oxine labeled WBC
- krypton-81m
- nitrogen-13
- oxygen-15
- phosphorus-32
- selenium-75
-
technetium agents
- Tc-99m DMSA
- Tc-99m DTPA
- Tc-99m DTPA aerosol
- Tc-99m HMPAO
- Tc-99m HMPAO labeled WBC
- Tc-99m MAA
- Tc-99m MAG3
- Tc-99m MDP
- Tc-99m mercaptoacetyltriglycine
- Tc-99m pertechnetate
- Tc-99m labeled RBC
- Tc-99m sestamibi
- Tc-99m sulfur colloid
- Tc-99m sulfur colloid (oral)
- thallium-201 chloride
- xenon agents
- in vivo therapeutic agents
- pharmaceuticals used in nuclear medicine
-
emerging methods in medical imaging
- radiography
- phase-contrast imaging
- CT
- deep-learning reconstruction
- photon counting CT
- virtual non-contrast imaging
- ultrasound
- magnetomotive ultrasound (MMUS)
- superb microvascular imaging
- ultrafast Doppler imaging
- ultrasound localization microscopy
- MRI
- nuclear medicine
- total body PET system
- immuno-PET
- miscellaneous
- radiography