Twinkling artifact

Last revised by Joshua Yap on 19 Aug 2022

Twinkling artifact is seen with color flow Doppler ultrasound 1. It occurs as a focus of alternating colors on Doppler signal behind a reflective object (such as a calculus), which gives the appearance of turbulent blood flow 2. It appears with or without an associated color comet tail artifact 3.

The underlying mechanism of this artifact is thought to be a result of inherent noise within the ultrasound scanner, specifically phase (a.k.a. clock) jitter within the Doppler electronics 10.

Twinkling artifact is more sensitive for detection of small stones (e.g. urolithiasis, cholelithiasis) than is acoustic shadowing. It is most pronounced when the reflecting surface is rough and highly dependent on machine setting:

  • when the focal zone is located below a rough reflecting surface, the twinkling artifact becomes more obvious than when it is above it 4
  • decreased pulse repetition frequency facilitates better visualization of the artifact 8,9

The presence of renal twinkling artifact on sonography has a high positive predictive value (78%) for the presence of nephrolithiasis on unenhanced CT 10.

History and etymology

The first description of twinkling artifact was by the French radiologist Alain Rahmouni (1957-2018) 7 et al. in 1996 5.

See also

ADVERTISEMENT: Supporters see fewer/no ads

Cases and figures

  • Case 1
    Drag here to reorder.
  • Case 2
    Drag here to reorder.
  • Case 3
    Drag here to reorder.
  • Case 4: from a foreign body
    Drag here to reorder.
  • Case 5: medullary nephrocalcinosis
    Drag here to reorder.
  • Case 6: vesicoureteric junction stone
    Drag here to reorder.
  • Case 7: vesicoureteric junction calculus
    Drag here to reorder.
  • Case 8: metal foreign body in arm
    Drag here to reorder.
  • Case 9: without color comet tail
    Drag here to reorder.
  • Case 10
    Drag here to reorder.