GOOD NEWS: We have fixed the DICOM uploading problem. New cases should work fine. More info radiopaedia.org/chat

Endoleak: type Ia following EVAS

Case contributed by Rish Ratnalingam
Diagnosis certain

Presentation

Acute, severe abdominal pain. EVAS device inserted 3 years ago.

Patient Data

Age: 85 years
Gender: Male

Imaging at 1 month post EVAS

ct

Routine surveillance imaging post endovascular aneurysm sealing system (EVAS) insertion demonstrates satisfactory positioning on the endovascular device. The proximal margins of the stents lie at the level of the right renal artery. There is a chimney graft extending from the left renal artery into the aorta. High attenuation polymer noted within the endobags.

3 yr post EVAS

ct

Both limbs of the device have migrated distally since the time of insertion resulting in a type Ia endoleak which is seen outlining the anterior margins of the endobags. 

There are concentric rings of high attenuation within the aneurysm sac, notable on the non-contrast images (dense rim sign) suggestive of acute hematoma and raising concern for impending rupture.

Case Discussion

Findings demonstrate a type Ia endoleak due to distal migration of the EVAS. The patient was not a candidate for an open repair and device removal and died shortly afterwards from a ruptured abdominal aortic aneurysm.