Presentation
US probe regularly used for biopsies.
From the case:
Signal dropout on US probe due to dead element
{"study":{"id":100753,"case_slug":"signal-dropout-on-us-probe-due-to-dead-element","modality":"Ultrasound","series":[{"id":54084665,"content_type":"image/jpeg","frames":[{"id":54084665,"width":1223,"height":792,"current":true}],"annotations":[],"encodings":{"original_dicom":null,"video":null,"thumbnailed_files":[{"tiny":"IMAGE_001-001_tiny.jpeg","jumbo":"IMAGE_001-001_jumbo.jpeg","small":"IMAGE_001-001_small.jpeg","thumb":"IMAGE_001-001_thumb.jpeg","medium":"IMAGE_001-001_medium.jpeg","gallery":"IMAGE_001-001_gallery.jpeg","big_gallery":"IMAGE_001-001_big_gallery.jpeg","original":"IMAGE_001-001.jpg"}]},"perspective":null,"specifics":null},{"id":54084666,"content_type":"image/jpeg","frames":[{"id":54084666,"width":1223,"height":792,"current":true}],"annotations":[],"encodings":{"original_dicom":null,"video":null,"thumbnailed_files":[{"tiny":"IMAGE_001-002_tiny.jpeg","jumbo":"IMAGE_001-002_jumbo.jpeg","small":"IMAGE_001-002_small.jpeg","thumb":"IMAGE_001-002_thumb.jpeg","medium":"IMAGE_001-002_medium.jpeg","gallery":"IMAGE_001-002_gallery.jpeg","big_gallery":"IMAGE_001-002_big_gallery.jpeg","original":"IMAGE_001-002.jpg"}]},"perspective":null,"specifics":null},{"id":54084667,"content_type":"image/jpeg","frames":[{"id":54084667,"width":1223,"height":792,"current":true}],"annotations":[],"encodings":{"original_dicom":null,"video":null,"thumbnailed_files":[{"tiny":"IMAGE_001-003_tiny.jpeg","jumbo":"IMAGE_001-003_jumbo.jpeg","small":"IMAGE_001-003_small.jpeg","thumb":"IMAGE_001-003_thumb.jpeg","medium":"IMAGE_001-003_medium.jpeg","gallery":"IMAGE_001-003_gallery.jpeg","big_gallery":"IMAGE_001-003_big_gallery.jpeg","original":"IMAGE_001-003.jpg"}]},"perspective":"Oblique","specifics":null}],"caption":{"us":null,"gb":null},"findings":{"us":"","gb":""},"contributor_login":"balintbotz","case_id":85201,"case_key_image_id":54084666,"case_featured_at":null},"current_user":null,"access":{"can_edit":false,"can_download":true,"can_feature":false,"can_pin":true}}
Routine inspection of in-air reverberation pattern reveals a tiny defect of the scan field, better appreciated with a reduced depth, and increased B-mode gain. Note that the artifact is relatively obscure, as a barely discernible band of reduced echogenicity during normal use.
Case Discussion
The ultrasound probe defect was caused by a barely visible puncture of the probe lens, resulting in the characteristic dropout artifact due to the dead piezo element.