Image intensifiers (II) are used to convert low energy x-radiation into visible light images. Frequently the detector portion of an x-ray C-arm used in operating theatres, the image intensifier has a low scatter input portion made of low absorption substances such as titanium or aluminium 1,2. Image intensifiers are several thousand times more sensitive compared to standard 400-speed screen-film combinations, and in practice can produce images using several thousand times less radiation 3,4.

The biggest advantage of image intensifiers in medical imaging is the synergy of high detector efficiency and high conversion efficiency to effectively utilise fluoroscopy while adhering to the radiation protection principle of dose optimisation.

Overview

II Components

After the x-ray beam emerges from the patient, it enters the image intensifier tube through the input window and is partially absorbed by the fluorescent input screen (entrance phosphor) creating a number of light photons 5.

The light photons strike the photocathode of the input screen and are absorbed by photoelectric interactions, causing it to emit photoelectrons (via the photoelectric effect) 5.

The electrons are accelerated towards the output fluorescent screen by an electric field produced between the photocathode and anode. Focussing and distortion minimisation is accomplished by the focussing electrodes 5.

The electrons hit the output phosphor and cause large numbers of light photons to be produced, which subsequently may be captured by various imaging devices 3,4.

Gain and conversion factor

The output of the image intensifier can be evaluated by brightness gain and conversion factor. Brightness gain is the product of minification gain and flux gain (also known as electron gain). Minification gain is the ratio of the input area to output area of phosphor. It makes the image brighter but does not improve the quality and contrast of the image. Meanwhile, flux gain is the number of photons generated at the output phosphor, comparing with photons generated at input phosphor 5.

Brightness gain from minification does not improve the statistical quality of the image because same number of photons is reaching output phosphor whether the input screen is larger or smaller. Thus, the total light output remains the same although minification gain results in increased brightness in the output screen 6.

Conversion factor is defined as output luminance of the image intensifier divided by entrance exposure rate, measuring the efficacy of conversion from x-rays to light. It has units of candela per square metre per milliroentgen per second ([cd/m2]/[mR/sec]) 5.

Vignetting

Vignetting refers to the reduced brightness observed at the peripheries of images captured using II. Often, anti-vignetting functions will be applied automatically to correct the image. Factors which contribute to this phenomenon include:

  • The curved input surface mean that the periphery of the input surface is further from the source and thus, due to the inverse square law, will receive a lower dose.

  • There is less scattered light in the peripheries of the output phosphor.

Distortion

II's are prone to two main types of distortion:

  • Pincushion distortion

    • Pincushion distortion describes stretching of the periphery of the image compared to the centre. It occurs because the input surface is curved, whilst the output surface is flat.

  • S distortion

    • S distortion describes warping of the fluoroscopic image in an 'S' pattern. It occurs due to the influence of magnetic fields outside of the II on electrons within the II. The addition of mu-metal to the housing unit of the II attempts to reduce this.

  • -<p><strong>Image intensifiers</strong>&nbsp;(<strong>II</strong>) are used to convert low energy <a href="/articles/x-rays-1">x-radiation</a> into visible light images. Frequently the detector portion of an x-ray <a href="/articles/c-arm">C-arm</a> used in operating theatres, the image intensifier has a low scatter input portion made of low absorption substances such as titanium or aluminium <sup>1,2</sup>. Image intensifiers are several thousand times more sensitive compared to standard 400-speed screen-film combinations, and in practice can produce images using several thousand times less radiation <sup>3,4</sup>.</p><p>The biggest advantage of image intensifiers in medical imaging is the synergy of high detector efficiency and high conversion efficiency to effectively utilise fluoroscopy while adhering to the <a href="/articles/radiation-protection">radiation protection</a>&nbsp;principle of dose optimisation.</p><h4>Overview</h4><p>After the x-ray beam emerges from the patient, it enters the image intensifier tube through the input window and is partially absorbed by the fluorescent input screen (<a href="/articles/entrance-phosphor">entrance phosphor</a>) creating a number of light photons <sup>5</sup>.</p><p>The light photons strike the <a href="/articles/photocathode">photocathode</a> of the input screen and are absorbed by photoelectric interactions, causing it to emit photoelectrons (via the <a href="/articles/photoelectric-effect">photoelectric effect</a>) <sup>5</sup>.</p><p>The <a href="/articles/electron">electrons</a> are accelerated towards the output fluorescent screen by an electric field produced between the photocathode and anode. Focussing and distortion minimisation is accomplished by the focussing electrodes <sup>5</sup>.</p><p>The electrons hit the <a href="/articles/output-phosphor">output phosphor</a>&nbsp;and cause large numbers of light photons to be produced, which subsequently may be captured by various imaging devices <sup>3,4</sup>.</p><p>The output of the image intensifier can be evaluated by brightness gain and conversion factor. Brightness gain is the product of minification gain and flux gain (also known as electron gain). Minification gain is the ratio of the input area to output area of phosphor. It makes the image brighter but does not improve the quality and contrast of the image. Meanwhile, flux gain is the number of photons generated at the output phosphor, comparing with photons generated at input phosphor <sup>5</sup>.</p><p>Brightness gain from minification does not improve the statistical quality of the image because same number of photons is reaching output phosphor whether the input screen is larger or smaller. Thus, the total light output remains the same although minification gain results in increased brightness in the output screen <sup>6</sup>.</p><p>Conversion factor is defined as output luminance of the image intensifier divided by entrance exposure rate, measuring the efficacy of conversion from x-rays to light. It has units of candela per square metre per milliroentgen per second ([cd/m<sup>2</sup>]/[mR/sec]) <sup>5</sup>.</p>
  • +<h4>Overview</h4><p><strong>Image intensifiers</strong>&nbsp;(<strong>II</strong>) are used to convert low energy <a href="/articles/x-rays-1">x-radiation</a> into visible light images. Frequently the detector portion of an x-ray <a href="/articles/c-arm">C-arm</a> used in operating theatres, the image intensifier has a low scatter input portion made of low absorption substances such as titanium or aluminium <sup>1,2</sup>. Image intensifiers are several thousand times more sensitive compared to standard 400-speed screen-film combinations, and in practice can produce images using several thousand times less radiation <sup>3,4</sup>.</p><p>The biggest advantage of image intensifiers in medical imaging is the synergy of high detector efficiency and high conversion efficiency to effectively utilise fluoroscopy while adhering to the <a href="/articles/radiation-protection">radiation protection</a>&nbsp;principle of dose optimisation.</p><h6>II Components</h6><p>After the x-ray beam emerges from the patient, it enters the image intensifier tube through the <strong>input window </strong>and is partially absorbed by the fluorescent <strong>input screen </strong>(<a href="/articles/entrance-phosphor">entrance phosphor</a>) creating a number of light photons <sup>5</sup>.</p><p>The light photons strike the <a href="/articles/photocathode"><strong>photocathode</strong></a> of the input screen and are absorbed by photoelectric interactions, causing it to emit photoelectrons (via the <a href="/articles/photoelectric-effect">photoelectric effect</a>) <sup>5</sup>.</p><p>The <a href="/articles/electron">electrons</a> are accelerated towards the output fluorescent screen by an electric field produced between the photocathode and <strong>anode</strong>. Focussing and distortion minimisation is accomplished by the <strong>focussing electrodes </strong><sup>5</sup>.</p><p>The electrons hit the <a href="/articles/output-phosphor"><strong>output phosphor</strong></a>&nbsp;and cause large numbers of light photons to be produced, which subsequently may be captured by various <strong>imaging devices </strong><sup>3,4</sup>.</p><h6>Gain and conversion factor</h6><p>The output of the image intensifier can be evaluated by <strong>brightness gain</strong> and <strong>conversion factor</strong>. Brightness gain is the product of <strong>minification gain </strong>and <strong>flux gain </strong>(also known as electron gain). Minification gain is the ratio of the input area to output area of phosphor. It makes the image brighter but does not improve the quality and contrast of the image. Meanwhile, flux gain is the number of photons generated at the output phosphor, comparing with photons generated at input phosphor <sup>5</sup>.</p><p>Brightness gain from minification does not improve the statistical quality of the image because same number of photons is reaching output phosphor whether the input screen is larger or smaller. Thus, the total light output remains the same although minification gain results in increased brightness in the output screen <sup>6</sup>.</p><p>Conversion factor is defined as output luminance of the image intensifier divided by entrance exposure rate, measuring the efficacy of conversion from x-rays to light. It has units of candela per square metre per milliroentgen per second ([cd/m<sup>2</sup>]/[mR/sec]) <sup>5</sup>.</p><h6>Vignetting</h6><p><strong>Vignetting </strong>refers to the reduced brightness observed at the peripheries of images captured using II. Often, anti-vignetting functions will be applied automatically to correct the image. Factors which contribute to this phenomenon include:</p><ul>
  • +<li><p>The curved input surface mean that the periphery of the input surface is further from the source and thus, due to the inverse square law, will receive a lower dose.</p></li>
  • +<li><p>There is less scattered light in the peripheries of the output phosphor.</p></li>
  • +</ul><h6>Distortion</h6><p>II's are prone to two main types of distortion:</p><ul>
  • +<li>
  • +<p><strong>Pincushion distortion</strong></p>
  • +<ul><li><p>Pincushion distortion describes stretching of the periphery of the image compared to the centre. It occurs because the input surface is curved, whilst the output surface is flat. </p></li></ul>
  • +</li>
  • +<li>
  • +<p><strong>S distortion</strong></p>
  • +<ul><li><p>S distortion describes warping of the fluoroscopic image in an 'S' pattern. It occurs due to the influence of magnetic fields outside of the II on electrons within the II. The addition of mu-metal to the housing unit of the II attempts to reduce this.</p></li></ul>
  • +</li>
  • +</ul>

ADVERTISEMENT: Supporters see fewer/no ads

Updating… Please wait.

 Unable to process the form. Check for errors and try again.

 Thank you for updating your details.